Ultrasonic clamp-on flowmeter Proline Prosonic Flow P 500 – noninvasive and easy to install – no process interruption – maintenance-free operation up to 550 °C
Specs at a glance
Max. measurement error
Volume flow: ±3% o.r. for DN 15 ±2% o.r. for DN 25 to DN 200 ±2% o.r. above DN 200
Measuring range
0 to 15 m/s (0 to 50 ft/s)
Medium temperature range
DN 15 to 65 (½to 2½"): –40 to +150°C (–40 +302 °F) DN 50 to 4000 (2 to 160"): –40 to +170°C (–40 to +338 °F) DN 50 to 600 (2" to 24"): 150°C to 550°C (+302 to +1022 °F)
Prosonic Flow P 500 is the ultrasonic clamp-on flowmeter of choice for safe and maintenance-free measurement of conductive and non-conductive liquids, even in challenging applications. It measures flow noninvasively and bidirectionally, independent of pressure, density and conductivity, and without process interruption.
FlowDC makes mounting as flexible as never before. The specified accuracy is maintained at an inlet run of only 2 x DN, even in the case of disturbed flow profiles.
Special coupling pads make this ultrasonic clamp-on system maintenance-free during the entire product life cycle.
In addition, Prosonic Flow P 500 is available with:
high-temperature sensors for measurement up to 550 °C (1022 °F)
a Petroleum Application Package (PAP), calculating the standard volume for liquid hydrocarbons
Heartbeat Technology, providing continuous process transparency. This way, build-up can be detected before it affects operation
Benefits
Flexible mounting – FlowDC reduces inlet runs to a minimum
High safety standards – Prosonic Flow P 500 is developed according to IEC 61508 functional safety requirements and holds international approvals for hazardous area, Zone 1 and Class I, Division 1
Reliable measurement performance – proven sensors in combination with maintenance-free mounting system deliver long-term stable signals, also in high-temperature applications (≤550 °C/1022 °F)
User-friendly operation – the web server offers full remote access, while the device’s touch display offers convenient on-site interaction
Transparent process insights – Heartbeat Technology with its integrated diagnostics, verification and monitoring functions enable compliance and process safety at all times
Low capital investment – cost-effectiveness increases with the pipe diameter (up to DN 4000/160")
Videos
Smart instrumentation through Heartbeat Technology
You want to increase plant performance and reduce costs? Our smart instrumentation portfolio with Heartbeat Technology offers in-depth device insight for more cost-efficiency and operational safety.
Schluchseewerk AG wanted to record and optimize the efficiency of its turbines at the pumped storage plant in Wehr. Endress+Hauser offers the optimum solution with its ultrasonic clamp-on flowmeter.
With Netilion Network Insights, Oberzent's water masters now see all relevant data points of their water system on a cloud-based dashboard without having to drive to the plants.
GPsol, together with Endress+Hauser, developed a system that individually measures each component of extraction wells and provides as much information as possible about the current flow.
Clamp-on flowmeter for limited spaces in process industries with up to 3 I/Os. Constant accuracy even when mounted with short inlet run thanks to FlowDC. Bidirectional measurement of various fluids, e.g. liquid hydrocarbons and chemicals.
Sensor features
High safety standards – SIL by design, international hazardous area approvals. Long-term stable signal – maintenance-free permanent mounting from outside with coupling pads. Reliable measurement on various pipe materials – sensor for GRP and plastic pipes available. Mounting without process interruption. Wide nominal diameter range: DN 15 to 4000 (½ to 160"). Medium temperature: –40 to +550 °C (–40 to +1022 °F).
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology. Remote version with up to 3 I/Os. Backlit display with touch control and WLAN access. Standard cable between sensor and transmitter.
Nominal diameter range
Dual channel, 1 or 2 paths : DN 15 to 4000 (1/2 to 160")
Volume flow: ±3% o.r. for DN 15 ±2% o.r. for DN 25 to DN 200 ±2% o.r. above DN 200
Measuring range
0 to 15 m/s (0 to 50 ft/s)
Max. process pressure
N/A
Medium temperature range
DN 15 to 65 (½to 2½"): –40 to +150°C (–40 +302 °F) DN 50 to 4000 (2 to 160"): –40 to +170°C (–40 to +338 °F) DN 50 to 600 (2" to 24"): 150°C to 550°C (+302 to +1022 °F)
Ambient temperature range
–40 to +60 °C (–40 to +140 °F) –50 to +60 °C (–58 to +140 °F) optional
Sensor housing material
N/A
Transmitter housing material
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L
Degree of protection
Sensor remote version: IP68 type 6P Transmitter remote version: IP66/67, Type 4X enclosure
Display/Operation
4-line backlit display with touch control(operation from outside) Optional: WLAN Configuration via local display and operating tools possible
Outputs
3 outputs: 4-20 mA HART (active/passive) 4-20 mA (active/passive) Pulse/frequency/switch output (active/passive) Double pulse output (active/passive) Relay output
Inputs
Status input 4-20 mA input
Digital communication
HART, Modbus RS485
Power supply
DC 24 V AC 100 to 230 V AC 100 to 230 V / DC 24 V (non-hazardous area)
Hazardous area approvals
ATEX, FM/CSA, EAC, UK Ex, KC
Other approvals and certificates
Other approvals and certificates
Product safety
CE, C-tick, EAC-marking
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
Metrological approvals and certificates
Flowmeter verification for all frequencies except 0.3 MHz, reference line size: 5 MHz DN 50, all other frequencies DN 100 Verification performed on accredited calibration facilities (acc. to ISO/IEC 17025) Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a
WirelessHART and/or Bluetooth® communication interface, Operating frequency: 2.4 GHz (ISM band), Transmission range under reference conditions: WirelessHART communication - up to 200 m (656 ft), Bluetooth communication - up to 40 m (131 ft)
WirelessHART and/or Bluetooth® communication interface, Operating frequency: 2.4 GHz (ISM band), Transmission range under reference conditions: WirelessHART communication - up to 200 m (656 ft), Bluetooth communication - up to 40 m (131 ft)